3.10 AC & DC Current Measurement
Connect the black test lead to the COM terminal and the red test lead
to the μA mA or 20A terminal depending on the magnitude of current
to be measured.
Set the rotary switch to either
A
or
A
and the required range.
Taking all necessary safety precautions connect the test leads in
series with the circuit being measured and read the measured current
from the display.
3.11 Resistance Measurement
Connect the black test lead to the COM terminal and the red test lead
to the
%HzV: terminal.
Set the rotary switch to : and the required range.
Taking all necessary safety precautions connect the test leads to the
circuit being measured and read the measured resistance from the
display.
3.12 Frequency Measurement
Connect the black test lead to the COM terminal and the red test lead
to the
%HzV: terminal.
Set the rotary switch to Hz and the required range.
Taking all necessary safety precautions connect the test leads to the
circuit being measured and read the measured frequency from the
display.
3.13 Duty Cycle Measurement
Connect the black test lead to the COM terminal and the red test lead
to the
%HzV: terminal.
Set the rotary switch to DUTY%.
Taking all necessary safety precautions connect the test leads to the
circuit being measured and read the measured % duty cycle from the
display.
MM84 instructions.indd 12 03/07/2012 15:48:30
7
3.14 Continuity Testing
Connect the black test lead to the COM terminal and the red test
lead to the
%HzV: terminal.
Set the rotary switch to
.
Taking all necessary safety precautions connect the test leads to
the circuit being tested.
The buzzer will sound if the resistance is <150:.
3.15 Diode Testing
If the diode to be tested is in circuit be sure the circuit power is
switched off.
Connect the black test lead to the COM terminal and the red test
lead to the
%HzV: terminal.
Set the rotary switch to
.
Taking all necessary safety precautions connect the test leads to
the diode being tested.
If the diode is good a forward bias will give a display reading of
around 0.6V (silicon diode) and a reverse bias will give a display of
OL. If the diode is shorted or open circuit the display will indicate
approx 0V or OL respectively for both forward and reverse bias.
3.16 Logic Testing
Connect the black test lead to the COM terminal and the red test
lead to the
%HzV: terminal.
Set the rotary switch to
LOGIC. The LCD will display LOGIC.
Taking all necessary safety precautions connect the test leads to
the circuit being tested.
A logic 0 (Lo) will be indicated by a
on the display.
A logic 1 (Hi) will be indicated by a
on the display and the MM84
buzzer will sound.
MM84 instructions.indd 13 03/07/2012 15:48:30
8
4. MAINTENANCE
4.1 Battery Replacement
To avoid shock or injury, disconnect the multimeter from any
external circuits or components and remove the test leads before
proceeding.
The battery is replaced by removing the rear casing of the MM84.
The rear casing is secured to the front by 3 screws and two internal
snaps at the LCD end.
Remove the rear casing by removing the three screws from the rear
casing. Gently lift the input terminal end of the rear casing until it
unsnaps from the casing top.
Fit a new 9V, NEDA 1604, JIS006P, IEC 6F22 battery observing
correct polarity.
NOTE: It is essential to transfer the protective insulating sleeve from
the old battery to the new battery.
Replace the rear casing by snapping it into position and re-install
the three screws.
4.2 Fuse Replacement
To avoid shock, injury or damage to the multimeter, disconnect
the multimeter from any external circuits or components and remove
the test leads before proceeding.
Replace only with the fuse specifi ed.
The fuses are replaced by removing the rear casing of the MM84.
Follow the procedure of section 4.1 (Battery Replacement) to
remove the rear casing.
MM84 instructions.indd 14 03/07/2012 15:48:31
9
Replace F1 only with the original type F500mA/600V, fast acting
ceramic fuse and replace F2 only with the original type F20A/600V
High Energy, fast acting ceramic fuse.
Replace the rear casing and screws.
4.3 Test Lead Replacement
If the test leads become damaged they should be replaced.
The replacement test leads must have the same (or better)
overvoltage category rating as the TL16 leads supplied.
4.4 Calibration
To maintain the integrity of measurements made using your
instrument, Martindale Electric recommends that it is returned
at least once a year to an approved Calibration Laboratory for
recalibration and certifi cation.
Martindale Electric is pleased to offer you this service. Please
contact our Service Department for details.
Email: service@martindale-electric.co.uk
Tel: 01923 650660
4.5 Cleaning
The unit may be cleaned using a soft dry cloth. Do not use moisture,
abrasives, solvents, or detergents, which can be conductive.
4.6 Repair & Service
There are no user serviceable parts in this unit other than those that
may be described in section 4. Return to Martindale Electric if faulty.
Our service department will quote promptly to repair any fault that
occurs outside the guarantee period.
Before the unit is returned, please ensure that you have checked the
unit, fuses, test leads and battery.